Effects of Sterilization on Cellobiose Dehydrogenase and Glucose Oxidase Based Glucose Biosensors

Résumé

Research on implantable glucose biosensors is driven by the need for innovative medical devices for continuous glucose monitoring in patients with diabetes mellitus. However, biosensor sterilization is a step that is widely omitted during the process of innovation. To compare the effects of gamma irradiation and chemical treatment with ethylene oxide (carbon microarray electrodes are fabricated, functionalized with glucose oxidizing enzymes (cellobiose dehydrogenase CDH or glucose oxidase GOx), and coated with a specifically designed zwitterionic polymer prior to the sterilization step. Cyclic voltammetry in the presence of 100 mm glucose of the biosensors before and after sterilization shows that gamma irradiation with a low radiation rate (25 kGy, 260 Gy h -1 ) does not induce a sensor performance loss, unlike the EtO treatment. In addition, no cytotoxic by-products are released after gamma sterilization. Based on these results obtained with both glucose oxidizing enzymes (CDH and GOx), gamma irradiation of the glucose biosensors with a low dose rate is preferable to exposure to EtO for biosensor terminal sterilization.
Fichier principal
Vignette du fichier
Advanced Sensor Research - 2024 - Bennett - Effects of Sterilization on Cellobiose Dehydrogenase and Glucose Oxidase Based.pdf (971.06 Ko) Télécharger le fichier
Origine Fichiers éditeurs autorisés sur une archive ouverte
licence

Dates et versions

hal-04777163 , version 1 (12-11-2024)

Licence

Identifiants

Citer

Richard Bennett, Arvind Rathore, Sébastien Gounel, Anna Lielpetere, Thomas M B Reichhart, et al.. Effects of Sterilization on Cellobiose Dehydrogenase and Glucose Oxidase Based Glucose Biosensors. Advanced Sensor Research, 2024, 3, pp.2400056. ⟨10.1002/adsr.202400056⟩. ⟨hal-04777163⟩
0 Consultations
0 Téléchargements

Altmetric

Partager

More